usually larger than the amount measured with the
unilateral cover test. The former is a result of the
sum of the fast and slow fusional vergence system,
whereas the latter is a measurement of the fast
fusion system. Patients who show strong vergence
adaptation and have high latent phorias will in-
crease the angle of deviation with prolonged, re-
peated alternate cover testing’, patients with large
phorias and weak slow vergence systems will not.
Thus, patients with robust vergence adaption are
less likely to be symptomatic, inasmuch as the slow
fusional vergence system eliminates the constant
demand on the fast fusional vergence system.
Burian and Smith 20 noted that both intermittent
exotropes and normals often show significant in-
crease in the measurement of their deviations when
fixation changes from 6 to 30 m (20 to 100
ft).
At
greater viewing distances there are fewer stereo-
scopic cues and less fusible detail, thus decreasing
reflex fusional vergence and subsequent slow fu-
sional vergence. This predictably results in a larger
deviation at a farther distance.
Prolonged Occlusion
Marlow 21 popularized a technique of prolonged
occlusion to uncover a latent heterophoria. He used
prolonged occlusion when traditional methods of
refractive or prismatic correction failed to provide
relief of asthenopia. The test consisted of prolonged
occlusion followed by. measurement of a phoria with
a Maddox rod. Care was taken not to allow any
opportunity for fusion during testing.
Beisbarth22 studied 29 patients with prolonged
occlusion (1 h to 9 days). Thirteen of 16 eyes with
preocciusion hyperdeviation showed a significant
increase in their deviation, 5 of 8 orthophoric pa-
tients demonstrated a postocclusion hyperphoria;
and the final 5 showed minimal changes with occlu-
sion. Beisbarth suggested that prismatic correction
based upon prolonged occlusion would result in
overcorrection of most vertical deviations. Pro-
longed occlusion eliminates vergence aftereffects,
resulting in an increase in both the number of
patients with a vertical deviation and in the
amount. However, robust vergence aftereffects
eliminate the load on the fast fusion system and
reduce asthenopia and diplopia. There is contro-
versy whether prolonged occlusion produces a true
deviation or just an exaggeration of Bell’s phenom-
enon. For a comprehensive review of this topic see
Amos and Rutstein.23
The important point to remember is that the
cover test does not measure the “true phoria.” Elim-
ination of all fusion-related impulses requires pro-
longed occlusion. The “latent deviation” identified
may in some cases be the elusive cause of astheno-
pia in patients. Thus, prolonged occlusion may be
used to identify binocular-induced symptoms by the
elimination of binocular fusion impulses and to
measure vergence aftereffects.
In 1952 Scobee24 occluded intermittent exotropes
of the divergence excess type for 1 h; he reported
that a significant percentage of them increased
their angular measurement at distance and near
after occlusion. Burian,25 based upon the results of
occlusion, classified the divergence excess type of
intermittent exotropia (DE) into two groups. One
group, which he called simulated DE, responded to
occlusion by increasing the angle of deviation at
near so that it approximated the distance deviation
(this represented 60% of the cases). The other 40%,
which were not affected by occlusion, were called
true DE.
ACA Ratio
Burian25 suggested that differentiation between
simulated and true DE was important because each
required a different surgical procedure. This finding
was confirmed by von Noorden26 though it has been
subsequently denied by other surgeons. This find-
ing with occlusion is important in understanding
the physiological mechanisms responsible for sen-
sory-motor functioning in DE. Before this finding,
most authorities reported that DE was associated
with a high ACA ratio. Using the distance near
formula, the minimum calculated ACA for a DE
patient, where the distance deviation is at least 10
greater than the near deviation, has to be at least
10/1.27
However, Scobee’s24 and Burian’s25 obser-’
vations with occlusion suggested that occlusion de-
creased the ACA ratio to approximately 6/1. Occlu-
sion should not have altered the true response ACA
ratio.
Cooper et al.,28 using an infrared measurement
system, measured accommodation and vergence si-
multaneously to determine objective, response
ACA’s. They demonstrated that response ACA’s in
patients with intermittent exotropia for both sim-
ulated and true DE were normal (mean
=
4.9) and
did not change with occlusion. It should be remem-
bered that all response ACA measurements, by the
nature of testing, use prolonged occlusion and
therefore eliminate vergence aftereffects as opposed
to clinical stimulus ACA measurements.
Cooper et al.28 postulated that the difference in
ACA ratios between objective, response ACA’s and
subjective, stimulus, distance-near ACA ratios in
patients with DE was due to the additive effects of
both vergence adaptation (slow fusional vergence)
and proximal convergence findings. They postu-
lated that most patients with simulated DE have a
robust slow vergence system, whereas most true DE
patients use excessive proximal vergence. Ob-
viously, some DE patients have mixed systems.
Kushner29 substantiated the findings of Cooper
et al. by studying gradient ACA’s in intermittent
exotropes. He reported that the majority (93%) had
normal gradient ACA’s. Only 7% of all DE patients
had a true high ACA. In those few DE patients
who had both high gradient and distance-near ACA
ratios, Kushner reported ,that surgical correction
resulted in a near esotropia. Thus gradient ACA’s,
near-far ACA’s, and occlusion testing should be
done on intermittent exotropes to determine the
Implications of Vergence Adaptation—Cooper. 303